Maintenance
141
Output Power
< 0 dBm
Match Measurement
Frequency Range
SA-2500A
SA-1700, SA-1700-P
SA-6000
780 – 2500 MHz
25 – 1700 MHz
25 – 6000 MHz
Frequency Resolution
SA-2500A
All other models
25 – 800 MHz
800 – 2500 MHz
2500 – 6000 MHz
50 kHz
25 kHz
50 kHz
100 kHz
Frequency Accuracy
± 150 kHz (75 ppm)
Number of Data Points
238 (default), 475, 949
Measurement Range and Resolution
Return Loss
VSWR
Range
Resolution
0.0 to –60.0 dB
0.1 dB
1.00 to 99.99
0.01
Measurement Uncertainty (after 1 display refresh)
*
Return Loss
Determined graphically from composite of
source VSWR, instrumentation error, and
directivity.
Refer to the figures on page 144, page 146,
VSWR
Calculated from Return Loss
*
Accuracy only guaranteed when using the Bird Calibration Combo
listed in this manual (See page 149).
Distance to Fault (DTF) Measurement
Mode
Internal
Resolution [R]