Bird Site Analyzer
152
Markers
Markers are used to show the exact value at one point
on the graph. In Measure Match and Cable Loss
modes, the marker displays the frequency and loss (or
VSWR). In Fault Location mode, the marker displays
the distance and loss.
Run / Hold
When
HOLD
is displayed, the unit makes a new
sweep every few seconds, and displays the latest
sweep.
When
RUN
is displayed, the unit will only make a sin-
gle sweep. The display will not update if the Site Ana-
lyzer is disconnected or if the cable is jiggled. This is
usually preferred.
Equations
Fault Location
Resolution
where C = speed of light (300 x 10
6
m / sec) or
(9.84252 x 10
8
feet / sec)
Maximum Distance
Fault Location Distance Tables
The tables on the next pages will help you select the
frequency span, measurable distance, and resolution
for fault location measurements. Select the table for
Resolution
C
freq span 2
×
-------------------------------
=
Max distance
resolution
data points 1
–
(
)
×
=
Содержание SITE ANALYZER SA-1700EX
Страница 2: ...This page is not blank...
Страница 8: ...Bird Site Analyzer vi...
Страница 9: ...vii...
Страница 30: ...Bird Site Analyzer 16...
Страница 142: ...Bird Site Analyzer 128...
Страница 148: ...Bird Site Analyzer 134...
Страница 159: ...Maintenance 145 Measured Return Loss dB...
Страница 161: ...Maintenance 147 SA 6000 Measurement Uncertainty Return Loss after one refresh Measured Return Loss dB Directivity...
Страница 167: ...Maintenance 153 the number of data points you wish measure 238 475 949...