Bird Site Analyzer
116
4.998 mW, a 0.04% error. Periodically rezero the sen-
sor. This applies to 5011 and 5012 power sensors.
Note:
For best results, make sure the sensor has
been connected to the Site Analyzer and the SA turned
on for at least 5 minutes. Do not apply RF power to the
sensor during calibration.
Press from
Measure
Power Mode
Press to
begin the
calibration
Fwd
0.00 mWatts
12.00
Offset
FullScale
0.00
dB
0.01 Watts
Zero
Calibrate
Recall
Setups
Measure Power Mode
Configuration Screen
This will take about 45
seconds to complete. Do
you wish to continue?
[Enter/Esc]