Key Components
|
39
•
When defining a filter set in a Read
step in a Gen5 procedure, selecting
“Hole” indicates the empty location in
the emission filter wheel. See page 43
for information on Read steps and
procedures.
The External Dispense Module
Applies only to Synergy HTX models with injectors.
The dispense module pumps fluid from the supply bottles to injector heads located
inside the instrument. Fluid is injected into one well at a time.
Figure 16: Dispense module components
•
Two 250-µL syringes draw fluid from the supply bottles.
•
Inlet tubes transport fluid from the supply vessels to the syringes. These tubes
are short pieces of opaque PTFE (Teflon) tubing connected to stainless-steel
probes on one end and threaded fittings on the other end.