Error Codes
|
159
Code
Description and Probable Causes:
4F0X
Fluorescence signal out of range (too low).
Too much light in chamber. Door not closed completely.
Ensure the Gen5 Filter Table accurately reflects the physical configuration of
the excitation and emission filter wheels.
Ensure the instrument case is completely installed and secured.
Sensitivity set too high. Try adjusting sensitivity setting.
Fluorescence standards dispensed to plate exceed value established by initial
standard values recorded.
Contamination within read chamber. Clean read chamber.
Verify there is no filter wavelength overlaps between excitation and emission
positions 2 and 3.
5000
5200
Carrier x-axis failed to home.
Shipping screw may be installed.
An object may be obstructing the path.
5001
5201
Carrier y-axis failed to home.
Y-axis rails are dusty or rusty. Dirt in the roller bearings is causing them to
jam.
An object may be obstructing the path.
5002
5202
Excitation filter wheel axis failed to home.
Filter wheel not inserted correctly.
Filter wheel obstructed.
Filter not clipped in.
Gear teeth of filter wheel binding with gear teeth of the motor. Remove filter
wheel, spin wheel by hand, and reinsert.
5003
5203
Emission filter wheel axis failed to home.
Filter wheel not inserted correctly.
Filter wheel obstructed.
Filter not clipped in.
Gear teeth of filter wheel binding with gear teeth of the motor. Remove filter
wheel, spin wheel by hand, and reinsert.
5006
5206
Probe z-axis failed to home.
Shipping bracket not removed.
5402
5403
Excitation filter wheel failed positional verify.
Emission filter wheel failed positional verify.
Ensure filter cartridge is fully inserted.
5700
5701
Carrier x-axis obstructed.
Carrier y-axis obstructed.
Axis may have hit probe z-axis.
Tip priming trough is not correctly inserted.
Synergy HTX Operator’s Manual
Содержание Synergy HTX
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Страница 2: ......
Страница 12: ...viii Preface Revision History Rev Date Changes A 7 2014 Initial release BioTek Instruments Inc ...
Страница 22: ...xviii Preface BioTek Instruments Inc ...
Страница 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 31: ...1 Unpack and Inspect the Reader 9 Figure 1 Unpacking the reader accessories box Synergy HTX Operator s Manual ...
Страница 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Страница 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Страница 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Страница 166: ......
Страница 174: ...150 Appendix A Specifications BioTek Instruments Inc ...