Chapter 3: Getting Started
•
Three-way valves switch the syringe flow from the inlet tubes to the outlet
tubes.
•
Outlet tubes transport fluid from the syringes into the instrument, through the
tubing ports on the Synergy HTX’s rear panel. The outlet tubes are opaque
PTFE tubes with threaded fittings on each end that are used to deliver fluid
from the syringes to the instrument.
Inside the Synergy HTX, two Teflon tubes transport fluid from the tubing ports on
the rear of the instrument to the two injectors. As shown below, both injectors are
positioned directly above the bottom fluorescence optical probe.
Syringe 1
Syringe 2
Bottom probe
Figure 17: Close-up view of the injectors inside the instrument
The tubing and injectors should be cleaned at least quarterly. See
Chapter 4, Preventive Maintenance for more information.
BioTek Instruments, Inc.