187
BM7Vet Pro
User’s Manual
a-
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) t
elephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broad
cast cannot be predicted theoretically with accuracy. To assess the electromagnetic environme
nt due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength outside the shielded location in which the EUT is used exceeds 3V/
m, the EUT should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocati
ng the EUT or using a shielded location with a higher RF shielding effectiveness and filter at
tenuation.
Note
For Type A Professional ME Equipment intended for use in domestic establishment instructions
for use includes a warning:
This ME equipment is intended for use by professional healthcare personnel only.
Warning
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the [ME
EQUIPMENT or ME SYSTEM], including cables specified by the manufacturer
Warning
Use of accessories and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation
Содержание BM7VET
Страница 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM7Vet Pro User s Manual BIGEMINY TRIGEMINY ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...