169
BM7Vet Pro
User’s Manual
NIBP Malfunction
- Confirm that the hose is normally connected
Are leaks from the hose
Yes
Connector of cuff
No
Abnormality In NIBP Measurements
The patient moved
While measuring
The measurements appear lower
No
The patient's skin
is extremely dry
No
Make sure the patient stay
still during diagnosis
Loosen the cuff during diagnosis
Tighten the cuff during diagnosis
Repair the NIBP B/D
No
Yes
Yes
Yes
Replace the hose of cuff
Repair the NIBP B/D