185
BM7Vet Pro
User’s Manual
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separ
ation distance (d) in meters (m) can be estimated using the equation applicable to the frequ
ency of the transmitter, where P is the maximum output power rating of the transmitter in w
atts (W) according to the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range ap
plies
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affec
ted by absorption and reflection from structures, objects, and people.
Immunity and Compliance Level
Immunity test
IEC 60601 Test Level
Actual Immunity Level
Compliance Level
Conducted RF
IEC 61000-4-6
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to
80 MHz
Radiated RF
IEC 61000-4-3
3 V/m, 80 MHz to 2.5
GHz
3 V/m, 80 MHz to 2.5
GHz
3 V/m, 80 MHz to
2.5 GHz