52
BM7Vet Pro
User’s Manual
Alarm Priority Alarm Sound
Alarm Color
Alarm Printer
Front Alarm
Lamp
Handle Alarm
Lamp
HIGH
-5
Every 2
seconds
Blinking
2 times
Blinking a
second
2 times
Blinking a
second
MEDIUM
-3
Every 2
seconds
Blinking
Every 2
seconds
Blinking
LOW
-2
Every 2
seconds
Blinking
Non-Blinking
MESSAGE
Non-Blinking
: Alarm sounds
: Waves are printed out
: Red color alarm indicator on the screen is blinked
: Yellow color alarm indicator on the screen is blinked
: Blue color alarm indicator on the screen is displayed
Audible Alarm
Alarm Priority
BIONET
IEC
HIGH
1 high tone every 5 seconds
5 consecutive beeps every 5 seconds
MEDIUM
1 high tone every 15 seconds
3 consecutive beeps every 15 seconds
LOW
1 low tone every 30 seconds
2 consecutive beeps every 30 seconds