118
BM7Vet Pro
User’s Manual
- Cause: Zero Required, Zero Error
- Solution: To clear, check airway adapter and clean if necessary. If this does not correct the
error, perform an adapter zero. If you must zero more than once, a possible hardware
error may exist.
* CO
2
OUT OF RANGE
- Cause: The value being calculated is greater than the upper CO
2
limit (150 mmHg)
- Solution: If error persists, perform a zero.
* CHECK AIRWAY ADAPTER
- Cause: Usually caused when the airway adapter is removed from the Capnostat or when there
is an optical blockage on the windows of the airway adapter. May also be caused
by failure to perform Capnostat zero when adapter type is changed.
- Solution: To clear, clean airway adapter if mucus or moisture is seen. If the adapter is clean,
perform a Capnostat zero.
Message
Status
Solution
MODULE OFF
It occurs when the equipment and
module are separated.
Verify module connections
Service request
Содержание BM7VET
Страница 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM7Vet Pro User s Manual BIGEMINY TRIGEMINY ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...