EPIC™
V Series
User Manual
pg. 12
5400430 Rev A
LASER
WARNING:
Always ensure that the protective laser eyewear is appropriate for the
laser wavelength.
•
Do not look directly into the beam or at specular reflections.
•
Never direct or point the beam at a person’s eyes.
•
Always place the system into STANDBY mode (by pressing the Control Button while in
READY mode) before exchanging Handpieces or disposable tips.
•
Toggle the ON/OFF switch (located on the rear of the console) to the OFF (O) position
before leaving unit unattended.
LASER
WARNING:
Do not open unit housing at any time. Danger from optical radiation may
exist.
LASER
WARNING:
Do not aim the laser at metallic or reflective surfaces, such as surgical
instruments or dental mirrors. If aimed directly at these surfaces the laser
beam will reflect and create a potential hazard.
CAUTION:
Be aware that the metal / plastic cannula on the tips may become hot during
use. Avoid contact of the cannula with any tissue.
3.3 SAFETY FEATURES
Energy Monitor
The energy monitor measures and verifies power output. Power deviations of more than ± 20%
from the selected value will cause the display to show the error message: “LASER CURRENT
HIGH/LOW”.
The laser console will not operate until the system first clears the error and then goes into
READY mode. If the error message persists, please contact BIOLASE Service at
1-800-321-6717.
System Monitor
The system monitors the emergency stop switch, remote key, wireless footswitch connection,
and output power. An error in any one of these will stop the system. The text display will indicate
the type of error. Operation will not resume until the error is cleared.