EPIC™
V Series
User Manual
pg. 7
5400430 Rev A
Figure 2.2:
Fiber
Optic Access Plug
Figure 2.3: Optical Access Port
2.7 SINGLE-USE TIPS
The tips are single-use accessories and are provided in three core diameters:
200μm, 300μm,
and 400μm, in different lengths (see Appendix C).
CAUTION:
Tips are single-use only
to avoid cross-contamination and are designed
to withstand a single sterilization cycle; they must be disposed of after
use in a biohazard medical waste Sharps container.
To connect the tip, insert it firmly into the distal end of the Handpiece as far as it will go, then
tighten by turning clockwise (Figure 2.7). Bend the metal cannula according to the specific
procedure requirements (Figure 2.10).
NOTE:
To provide proper laser operation, do not connect tips when the Handpiece is
disconnected.
2.8 SURGICAL HANDPIECE ASSEMBLY
► To connect the Handpiece to the fiber optic assembly, push the Handpiece on the fiber
shaft until it clicks on and is secured at connected position.
► To disconnect the Handpiece from fiber optic assembly:
•
Take the Handpiece body in one hand and the shaft in another
•
Push the two buttons on the Fiber Shaft
•
Pull the Handpiece with the ring to separate.
► Remove the fiber tip by twisting the tip counterclockwise.