EPIC™
V Series
User Manual
pg. 41
5400430 Rev A
Title
Message
Reason
Fix
Error 1
Thermistor Open
Thermistor Open
Call BIOLASE Service
Error 2
Thermistor Shorted
Thermistor Shorted
Error 3
Shutdown Temperature
System too hot
Allow 5-10 mins for laser to cool
down
Error 4
Laser Current High/ Low
Output is out of specs
Call BIOLASE Service
Error 5
FS shorted in Standby
FS is partially pressed or
is damaged
Press/Release FS or call Biolase
Service
Error 6
ON/OFF button Stuck
Key stuck
Press Front key
Error 7
Flash Corrupted
Memory Corrupted
Call BIOLASE Service
Error 8
No Fiber
Fiber not inserted
Plug in Trunk Fiber
Error 9
Lost Footswitch Communication Wireless Interference
Reposition console or FS to
improve communication
Error 10
Emergency Switch
E-Switch Pressed
Press E-Switch Again
Error 11
Remote Interlock
Remote interlock open
Check Remote Interlock closed
Error 12
Battery Critically Low
Battery Critically Low
Plug in DC supply
Warning 1
Temp High
System is hot
Allow 5-10 mins for laser to cool
down
Warning 2
Battery Low
Battery is low
Plug in DC supply
Warning 3
Battery Not Connected
Battery not connected
Plug in Battery
Warning 4
FS Battery Low
Battery on FS low
Replace FS battery
Alert 1
Wireless Not Paired
No wireless connect
Re-establish pairing (see Sec 4)
Alert 2
System must be in READY
mode to lase
System is not in READY
mode
Press the Control Button in any
procedure screen
Figure 11.1