EPIC™
V Series
User Manual
pg. 36
5400430 Rev A
8. MAINTENANCE
WARNING:
No modification of this equipment is allowed.
8.1 DAILY MAINTENANCE
Use the peel-off clear covers for the laser console supplied with the system. Use disinfectant to
wipe down the front panel and Handpiece holder of the EPIC V-Series system after each
procedure.
Do not use bleach or abrasive cleansers.
8.2 CLEANING AND STERILIZATION PROCEDURES
The contamination control suggested for the EPIC V-Series Surgical Handpiece and tips is the
steam sterilization method. However, before sterilization, the EPIC V-Series reusable
Handpiece should be carefully cleaned per the following procedure.
CAUTION:
Handpiece and tips must be cleaned and sterilized prior to initial use.
Tips are single-use only
to avoid cross-contamination and are designed to
withstand a single sterilization cycle; they must be disposed of after use in a
biohazard medical waste Sharps container.
Handpieces are reusable and must be cleaned and sterilized between
patients to avoid cross-contamination.
Cleaning and Disinfecting Instructions for the Surgical Handpiece, and the
Reusable Fiber Optic Cable
The cleaning process is intended to remove blood, protein and other potential contaminants
from the surfaces and crevices of reusable accessories. This process may also reduce the
quantity of particles, microorganisms and pathogens present. Cleaning should be performed
prior to sterilization and must be conducted only by qualified office personnel trained to perform
the procedure and handle the EPIC V-Series fiber optic delivery system.
Wear protective latex gloves when handling the contaminated delivery system.
To disinfect the fiber cable, wipe the entire cable, including the shaft, with an appropriate
disinfecting solution, such as Cavicide™ or a similar quaternary ammonium compound
product (containing 20% alcohol or less), and follow the manufacturer’s instructions. Avoid
getting any liquid or debris near the distal end of the fiber cable.