EPIC™
V Series
User Manual
pg. 30
5400430 Rev A
Tip Diameter (µm)
(Preset) Power (W)
Mode
400
1.4
CW
300
1.4
CW
200
Tip initiation not required when used for recommended procedures
Figure 7.2
● Touch the tip to the surface of the initiation block, without
activating the laser (don’t press down on the footswitch
(Figure 7.3).
Figure 7.3
● Press the footswitch to activate the laser, allowing the tip to
sink into the block. Pull the tip out when the metal cannula
touches the block, still firing until just before the tip is out of
the block (Figure 7.4).
Figure 7.4
● Press the footswitch to activate the laser into the air once, you
will see a white flash or the tip will glow (Figure 7.5).
Figure 7.5
● Repeat initiation process as needed to ensure the tip is initiated.
After tip initiation is completed, press the check mark to access the screen for the selected
procedure.
Figure 7.6
CAUTION:
If the laser console is in “READY” mode, the laser will fire if the footswitch is
activated.
LASER
WARNING:
Never point the laser at a person’s eyes. All persons present in the
operatory must wear protective eyewear when the laser is in operation