Muscle tissue closer to the skin surface may experience a higher absorption of heat;
carefully monitor skin temperature and reduce power as necessary.
•
Excessive fatty tissue is known to transmit heat without much attenuation; reduce power.
•
Different implant materials will respond differently to laser energy and heat; be aware of
any implants and their location; avoid direct exposure to laser energy or heat at the site of
the implant.
•
Do not apply ointments, creams, lotions or heating lotion patches at, or in close proximity
to, the site of treatment.
•
Do not apply therapies prior to treatment that could change body temperature, such as
ultrasound, ice/heat pack, electrical stimulation, or heating patches.
Using the Deep Tissue Handpiece
If holding the Handpiece in a constant location, the recommended initial power setting for
therapeutic effect is 4.0W CW delivered over 10 minutes of continuous treatment, with the
spacer set at a 30mm spot size. Always monitor patient response; adjust power and/or distance