KB ECO / KB ECO-UL (E6.2) 12/2021
Page 55/85
18.6 Removing the USB stick
Logging off the USB stick is not possible / required.
Be sure that no data recording is running (chap. 18.4).
After removing the USB stick, the controller is back in the same menu as before when connecting the USB
stick.
19. Self-test function
The self-test function enables an automated check of the proper chamber functioning as well as a targeted
and reliable fault analysis. In this case, the chamber successively undergoes various defined operating
states, which serves to determine reproducible characteristic values. These characteristic values provide
information on the performance and precision of the individual functional systems of the chamber (e.g.,
heating, refrigeration) of the chamber.
The results of the self-test are stored in the service recorder of the controller. You can export them using
the controller’s USB interface and send them to BINDER Service (use function “Export recorder data” to
USB stick, chap. 18). BINDER Service will evaluate the data using an analyzing tool.
In order to allow an optimum comparison of the determined characteristic values with the
reference characteristic values, the ambient temperature should be in the range of +22 °C
+/- 3 °C /
71.6 °F +/- 5.4 °F
.
The chamber shall be unloaded (empty with standard equipment).
19.1 Activating the self-test function
Required access level: “Admin”.
Path:
Normal display
Settings
Various
Self-test
Press the
OK button
to enable the settings.
Following, you can choose between the following settings:
•
“0” (function off)
•
“1” (self-test; duration: 12-18 hours)
Setting the self-test function
The current setting flashes. Use the
arrow buttons
to select the set-
ting “1” to activate the self-test function.
Confirm the setting with the
OK button
.
Self-test
Press the
Back button
several times to go back to
Normal display
.
In Normal display the activated Self-test function is shown. The “Collective alarm” icon is lit. The entered
setpoint values are inactive, the stored test program is running. With activated audible signal: The buzzer
sounds. Press the
OK button
to mute the buzzer.
Содержание 9020-0423
Страница 72: ...KB ECO KB ECO UL E6 2 12 2021 Page 72 85 26 7 Dimensions Dimensions size 240 mm...
Страница 73: ...KB ECO KB ECO UL E6 2 12 2021 Page 73 85 Dimensions size 720 mm...
Страница 74: ...KB ECO KB ECO UL E6 2 12 2021 Page 74 85 Dimensions size 1020 mm...
Страница 76: ...KB ECO KB ECO UL E6 2 12 2021 Page 76 85...
Страница 77: ...KB ECO KB ECO UL E6 2 12 2021 Page 77 85...
Страница 79: ...KB ECO KB ECO UL E6 2 12 2021 Page 79 85...