KB ECO / KB ECO-UL (E6.2) 12/2021
Page 38/85
11.2.3 Message and measures in the state of alarm
The state of alarm is indicated visually in Normal display. If the buzzer is enabled (chap. 14.2) there is an
additional audible alert. The heating turns off. As soon as the inner chamber temperature has cooled down
below the safety controller value, the heating is released and temperature control continues.
In Normal display a text message indicates the alarm cause. The “collective alarm” icon is lit. If the audible
arm is activated, the buzzer sounds. Press the
OK button
to mute the buzzer.
The alarm message “Safety controller” and the “Collective alarm” icon are displayed on the controller until
you press the
OK button
on the controller
and
the inner chamber temperature has cooled down below the
safety controller value.
•
If the inner chamber temperature has already cooled down below the safety controller value when press-
ing the
OK button
, the alarm message “Safety controller” and the “Collective alarm” icon are reset
together with the buzzer.
•
If the state of alarm is still active when pressing the
OK button
, i.e. the inner chamber temperature is
still above the safety controller value, first only the buzzer is reset. The alarm message “Safety controller”
and the “Collective alarm” icon will disappear as soon as the inner chamber temperature falls below the
safety controller value.
The alarm remains active until it is acknowledged on the controller and the inner temperature falls below
the entered safety controller setpoint. Then the heating is released again.
Normal display showing safety controller alarm (sample values)
Safety controller
Note:
When the safety controller class 3.1 had been activated you should disconnect the chamber from the power
supply and have an expert examine and rectify the cause of the fault.
11.2.4 Function check
Check the safety controller at appropriate intervals for its functionality. It is recommended that the author-
ized operating personnel should perform such a check, e.g., before starting a longer work procedure.
Содержание 9020-0423
Страница 72: ...KB ECO KB ECO UL E6 2 12 2021 Page 72 85 26 7 Dimensions Dimensions size 240 mm...
Страница 73: ...KB ECO KB ECO UL E6 2 12 2021 Page 73 85 Dimensions size 720 mm...
Страница 74: ...KB ECO KB ECO UL E6 2 12 2021 Page 74 85 Dimensions size 1020 mm...
Страница 76: ...KB ECO KB ECO UL E6 2 12 2021 Page 76 85...
Страница 77: ...KB ECO KB ECO UL E6 2 12 2021 Page 77 85...
Страница 79: ...KB ECO KB ECO UL E6 2 12 2021 Page 79 85...