KB ECO / KB ECO-UL (E6.2) 12/2021
Page 25/85
Functional controller keys
Icon
Signification
Function
Arrow-up button
•
Navigate between menus, submenus, other functions
•
In the setting menu: change setting, decrease value
Arrow-down button
•
Navigate between menus, submenus, other functions
•
In the setting menu: change setting, increase value
OK button
•
Select menu, submenu, function
•
In the setting menu: Confirm entry
Back button
Back to previous menu level
Standby button
no function
5.1 Menu structure of the controller and access levels
Starting from
Normal display
, navigate between the menus with the
arrow buttons
.
With the
OK button
you enter the setting of further subordinate menu functions.
With the
Back button
you go back to the previous function and finally back to Normal display.
The available functions depend on the current dependent on the current
authorization
“User”, “Admin” or
“Service”, for which the entry of a password may be required, depending on the setting.
You can set passwords for different access levels:
•
User:
The password enables access to the standard operating functions. Factory setting is 00 00 (no
password assigned).
•
Admin:
The password enables access to advanced controller functions and settings. Factory setting is
00 01.
•
Service:
The password enables access to all controller functions (for BINDER Service only).
As soon as a password has been assigned, access to the respective functions is blocked and only available
after entering the correct password.
Menu
Required access level
Functions
Setpoints
“User”
•
Temperature set-point setting
•
Setting the safety controller
•
Activating/deactivating operating mode “Idle mode”
Chamber info
Any user
•
Configuration display (setup information, controller hard-
ware and software, analog inputs)
•
Display of interface configuration (e.g. MAC address, IP
address)
Settings
“Admin”
•
General controller settings (date, time, menu language,
temperature unit, display brightness…)
•
Network settings
•
Setting the data logger storage interval
•
Setting the tolerance range limits and delay time for tol-
erance range alarm
•
Self-test function
•
Password changing for User and Admin
Содержание 9020-0423
Страница 72: ...KB ECO KB ECO UL E6 2 12 2021 Page 72 85 26 7 Dimensions Dimensions size 240 mm...
Страница 73: ...KB ECO KB ECO UL E6 2 12 2021 Page 73 85 Dimensions size 720 mm...
Страница 74: ...KB ECO KB ECO UL E6 2 12 2021 Page 74 85 Dimensions size 1020 mm...
Страница 76: ...KB ECO KB ECO UL E6 2 12 2021 Page 76 85...
Страница 77: ...KB ECO KB ECO UL E6 2 12 2021 Page 77 85...
Страница 79: ...KB ECO KB ECO UL E6 2 12 2021 Page 79 85...