FLYNET 100
5.2.1.2 Stamping management
Allows you to set the standard and the type of stamping.
STANDARDS
Not Legal
It allows deciding whether the scale is to be used for
legal purposes (e.g. commercial transactions with verification by the component
metrology
inspector), or for internal purposes. When enabled, it is not possible to change
sensitive data about the scale unless the hardware or software metric seal is
removed. If the indicator is approved, it is possible to display the data but not
change them. When one of the standards is applied, it is shown on the main scale
configuration page.
European
Type of compliance valid for the European Community
International International type of compliance
USA
Type of compliance valid for the USA (NTEP)
TYPE OF STAMPING
Hardware
The calibration button temporarily disables the operation in legal mode. To
temporarily disable stamping, enter the Scale Configuration page and press the
calibration button. When the indicator is stamped, the page has a red background,
once the seal is removed it turns white.
Software
The metrological password temporarily disables the operation in legal mode. To
temporarily disable stamping, enter the stamping management page and press the
icon
, then the page for entering the metrological password will
open.
5.2.1.3 Metrological information
Allows consulting the metrological information.
Serial number
Indicator serial number (protected by metrology)
Scale firmware
ID, Programme checksum release
Metrological modules Metrology software release and checksum
Stamping Status
Current status of the stamping and counter of metrological parameter
accesses