FLYNET 100
Problem
Cause
Remedy
- 01 -
Faulty converter
Incorrect or missing load
cell connection
Check extension cable, junction box and load
cells
- 01 -
Faulty converter
(*) Return signal from load
cell out of range
Check that the load cell is efficient and that it
works within its field of use
- 01 -
Faulty converter
(*) Board breakage
Check the conversion board with a load cell
simulator. If the test result is negative, replace
the converter board
- 01 -
Faulty converter
(**)
Digital
cells
not
responding
Check cell connections
- 02 -
Error in parameter
memory
Data
error
in
board
parameter memory
Check linearised sampling parameters
- 06 -
(**) Serial number error on
digital cell
Check the serial numbers and, if necessary,
carry out angle calibration
- 07 -
(**) Breakage of a cell on
multi-cell system
Replace the cell and carry out angle calibration
or angle calibration for the individual cell
- 08 -
(**) Scale board breakage
Replace the scale board and carry out angle
calibration or restore the indicator data (if they
have been saved before the breakage)
- 09 -
(**) First system switching
on
Carry out angle calibration
- 10 -
(**) At least one cell not
configured
Carry out angle calibration
- 11 -
(**) Digital cell power
supply problems
Check cables and power supply voltage
- 12 -
(**) Digital cell power
supply problems
Replace the load cell that signals the error
- 13 -
(**)
Cell
with
internal
temperature outside the
range (-40 ÷ 100°C)
Replace the cell that signals the error.
Carry out angle calibration for the replaced cell.
- 21 -
(**) Temporary digital cell
error
--