![Bilanciai D70ES Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/bilanciai/d70es/d70es_use-and-maintenance-manual_2750937076.webp)
1-76
D70ES
10.4.2 Weighing operation request from serial command
with transmission at end of operation
After loading the weight onto the weighing platform, send the storage
command and data request through the calculator:
MP<CR>
With an affirmative response the terminal stores the weight and transmits a
data string described in par. 10.5.
Therefore, the calculator responds with: ACK/NAK, in the same manner as
described in the previous paragraph.
10.4.3 Weighing request from key and transmission from
serial command
After loading the weight onto the weighing platform, wait for it to indicate that
the weight is stable, then press
. The terminal saves the data.
The calculator can request the weight data from the terminal through the
following command:
MP<CR>
The weighing terminal can respond with one of the strings described in par.
10.5; or
NP<CR><LF>
if no weight has been stored;
The data can be requested several times by simply repeating the
MP<CR>
command.
At the end of the operation the calculator must send the command to enable
the terminal for a new weighing operation:
MC<CR>