DIH 0300 B – USER GUIDE REANIBEX 300
56
9 DATA MANAGEMENT
9 DATA
MANAGEMENT
9.1
DESCRIPTION
The REANIBEX 300 can store the last 10 episodes of the equipment in its internal memory.
An episode consists of the continuous ECG signal and its associated events (lead change, alarms, etc.)
from the time the equipment is switched on or until the battery runs down. In addition, the biological
parameter (heart rate) is stored regularly. The episode will be stored even if the shutdown is uncontrolled,
i.e., even with empty or remove the battery during this episode.
When the equipment is working in Automated Defibrillator mode, the equipment and setting audio will be
recorded, as long as this option is activated in the Configuration.
Each of these uses is identified unequivocally in the internal memory by the date and time it was
started on.
The use will be registered in the internal memory even if the equipment switches off due to battery
removal or depletion.
In addition to this information, the equipment stores the last 100 tests performed in its internal memory,
including the type of test, the result and date the test was performed.
The REANIBEX 300 also indicates any incidents affecting the internal memory in charge of saving the
episodes, since the memory may be affected by the large number of writings and readings carried out.
If the internal memory does not have any problem storing the information, the icon
will appear. If
the equipment detects an error in the internal memory at any time, the icon
will appear. When a
large amount of episodes have been stored, the user will be informed that complete deletion of the
internal memory is recommended. The user will be shown the memory status with the icon
flashing.
Complete deletion of the internal memory is carried out in the Configuration mode. Refer to Section
"7.5.4
EPISODE DELETION"
for further details.
All the information registered in the internal memory of the equipment can be reviewed afterwards in a PC
by using the application for that purpose.
9.2
STORED EVENTS
All the events that are stored along with the ECG signal during registration of the intervention are shown
below.
The texts in cursive are replaced with the appropriate value or option.
EVENT
DESCRIPTION
SWITCHING ON
Switching on the equipment
MANUAL DEFIBRILLATOR MODE
Changes to Manual Defibrillator mode
Содержание REANIBEX 300
Страница 1: ...AED Manual Defibrillator REANIBEX 300 USER MANUAL DIH 0300 B REV C 2014 FEB...
Страница 2: ......
Страница 3: ...REANIBEX 300 USER GUIDE Review Date Approved by Signature C February 2014 Gorka Landaburu R D Manager...
Страница 4: ...Blank page...
Страница 6: ...Blank page...
Страница 8: ...Blank page...
Страница 12: ...DIH 0300 B USER GUIDE REANIBEX 300 iv TABLE OF CONTENTS...