4 Operation with ICE
LB 9510 Lumat
Page
60
of
79
71450BA2 Rev. 00, 12/2020
4.7.2
Direct data print-out
1.
If not opened already, open the respective file by selecting
[File] /
[Open data file].
2.
Select
[File] / [Print preview]
to get a preview of your print-out.
3.
Select
[File] / [Print]
or click on the printer icon in the left-hand navi-
gation bar to start printing the data.
4.7.3
Data Export via Report
IMPORTANT
Report
The report function is only functional with immunoassay measure-
ments.
1.
Click on
<Report>
. The
[Report form]
will open.
The report is displayed in its preset and assay-type specific layout.
The Main Report which is displayed by default contains
•
Date and time
•
Measurement file name
•
Protocol name and type
•
Measuring Time, Lot Number and Comment
•
Standards section with single and average reader values, nominal
concentrations and CVs
•
Standard concentration recalculation section with the nominal
concentrations, the recalculated concentrations, the respective de-
viations and the percentage in relation to the maximum
concentration and the curve fit data
•
Graphical display of the standard curve
•
Controls section with single and average reader values, calculated
concentrations and CVs
•
Quality control evaluation section with the controls seen in relation
to their expected ranges
•
Samples section with single and average reader values, calculated
concentrations and CVs
•
Statistics section with reader information, program version and
Operating system.