4 Operation with ICE
LB 9510 Lumat
Page
52
of
79
71450BA2 Rev. 00, 12/2020
[%CV
of ……]
Define the max. acceptable coefficient of var-
iation of the replicates of the samples and the
negative control, above which the results will
be flagged.
Use the settings as described below when
•
negative and positive controls are measured besides the samples:
[use negative control]
checked
[No. of positive controls]
1-3
[Replicates of Negative control]
Define the number
[Replicates of Positive control]
Define the number
[Low Cut Off]
i.e. the RLU value below which a result will be
flagged negative
Define the
[Offset] (AL)
: a fixed value
Define the
[Factor] (BL)
: a factor to determine
the negative control’s weight on the low cut
-
off.
Define the
[Lpos factor] (CL)
: a factor to de-
termine the positive control’s weight on the
low cut-off.
[High Cut Off]
i.e. the RLU value above which a result will be
flagged positive.
Define the
[Offset] (AH)
: a fixed value
Define the
[Factor] (BH)
: a factor to deter-
mine the negative control’s weight on the
high cut-off