LB 478 MPLM
9 Troubleshooting
56925-8BA2 Rev.00, 07/2019
83
9
Troubleshooting
9.1
Error Search
Problem
Cause
Measure
Master unit: Screen black; LEDs
are not illuminated
EVU does not work
u
Check power supply and
fuses
Extension module:
LEDs are not illuminated
Extension module not
clamped properly
u
Check cabling, contact so-
ckets
Extension module:
is not detected
Incorrect device ID
u
Check Device ID in the
menu
Identification
of the
extension module
No signal
Detector does not work
u
Check the functioning of
the detector
Count rate too low
Shield not opened or not
opened correctly
u
Check lock and ensure it is
in OPEN position
Incorrect focus of the effec-
tive radiation on the
detector
u
Correct and optimise the
alignment
Objects in the beam path
u
Offset irradiation level
Source at the end of its usa-
ble life span
u
Replace source
The display deviates
Defect in detector
u
Check detector
Count rate too low (see
above)
u
Check source age and irra-
diation level, replace
detector
Incorrect Calibration
u
Check calibration values
Detector is not detected (soft-
ware)
Terminals / wiring
u
Check terminal connec-
tion; check terminal
assignment
Damaged line
u
Check cable; examine with
measurement device.
Incorrect ID in the configura-
tion
u
Check ID of the detector
(see type plate on the de-
tector)
Touch panel does not respond
Error in operating system
u
Restart EVU
Buttons are missed when you
click
Incorrect screen calibration
u
Calibrate screen again
Содержание Duo Xpert LB 478 MPLM
Страница 2: ......
Страница 4: ......
Страница 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Страница 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Страница 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Страница 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Страница 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Страница 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Страница 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Страница 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Страница 109: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 03 05 2019...
Страница 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...
Страница 130: ...Technical Information EmulsionSENS Technische Information EmulsionSENS 56925TI1M Rev 00 07 2019...