7 Main Menu Device Setup
LB 478 MPLM
52
56925-8BA2 Rev.00, 07/2019
Input / Touch
Device Setup | Setup | System | Interfaces | Local Display | Input / Touch
1 Button
<Calibrate touchscreen>
Fig. 30
Input/ Touch
Calibrate touch screen
The calibration may only be carried out with direct skin contact. Take gloves or any
other protective equipment off your hands.
Fig. 31
System menu, Settings - Display (Calibrate touch display)
1.
Click on
<Calibrate touch screen>.
u
The calibration screen opens.
2.
Press the middle of the displayed cross with your finger.
u
If you take your finger off the cross again, the cross jumps to the top left cor-
ner.
3.
Repeat the process until the cross is no longer displayed and the calibration is
finished.
4.
Confirm the calibration by clicking on the white screen to go back to "In-
put/Touch"
5.
Execute a restart of the EVU after prompting.
u
The screen calibration was performed.
Содержание Duo Xpert LB 478 MPLM
Страница 2: ......
Страница 4: ......
Страница 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Страница 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Страница 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Страница 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Страница 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Страница 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Страница 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Страница 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Страница 109: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 03 05 2019...
Страница 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...
Страница 130: ...Technical Information EmulsionSENS Technische Information EmulsionSENS 56925TI1M Rev 00 07 2019...