12 Appendix
LB 478 MPLM
102
56925-8BA2 Rev.00, 07/2019
Calibration of detectors
NOTE
u
Each detector must be calibrated individually!
u
As part of the calibration procedure, the container must be emptied or filled,
time must be allowed for this.
For the detectors of a MPLM system, a two-point calibration is performed, i.e. first
the calibration of all detectors is carried out at density A and then at density B. In
individual cases, a 1-point calibration may suffice. The calibration of the detectors
is described in detail in the corresponding operating instructions, at this point only
assistance is given. Calibration is documented individually for each detector in the
form of a commissioning protocol (template see 12.2 Setup Protocol LB 478).
1.
Connect the HART communicator to detector #1
Note: LB480 Evaluation Units are passive, 24 V auxillary voltage in series to
4 - 20 mA output required.
2.
Set the basic parameters.
3.
Calibrate first density.
4.
Repeat Step 2 and 3 for the remaining detectors.
5.
Calibrate the second density at all detectors and trigger calibrate.
Содержание Duo Xpert LB 478 MPLM
Страница 2: ......
Страница 4: ......
Страница 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Страница 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Страница 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Страница 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Страница 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Страница 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Страница 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Страница 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Страница 109: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 03 05 2019...
Страница 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...
Страница 130: ...Technical Information EmulsionSENS Technische Information EmulsionSENS 56925TI1M Rev 00 07 2019...