7 Main Menu Device Setup
LB 478 MPLM
68
56925-8BA2 Rev.00, 07/2019
Perform calibration
j
DANGER
Danger to life from electric shock!
u
The repair may only be carried out by a qualified electrician.
u
Please adhere to the relevant safety regulations.
u
During servicing on the hardware as well as during wiring of the detector,
the measuring system, connected relay contacts and all inputs and outputs
must be de-energised.
In case of an electric shock, carry out first aid measures and immediately call an emer-
gency service.
1.
Connect the measuring lines of the current measuring device on the back of
the EVU to the analog output.
2.
Observe the terminal allocation in the document “Technical information”.
3.
Click on the button
<Calibration>
.
u
The device switches to test mode and a new window (Analog output calibra-
tion) opens.
u
The calibration point 4 mA is displayed and the current measuring instrument
shows a value.
4.
Enter the indicated value on the current measuring instrument in the input
box (Fig. 49, item 4).
5.
Click on the button
<Continue>
.
u
The calibration point 20 mA is displayed and the current measuring instrument
shows a value.
6.
Enter the indicated value on the current measuring instrument in the input
box (Fig. 49, item 5).
7.
Click on the button
<Continue>
.
u
A message appears “Calibration successful”.
8.
Click on the button
<Continue>
.
u
The calibration of the analog output is completed.
Содержание Duo Xpert LB 478 MPLM
Страница 2: ......
Страница 4: ......
Страница 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Страница 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Страница 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Страница 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Страница 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Страница 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Страница 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Страница 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Страница 109: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 03 05 2019...
Страница 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...
Страница 130: ...Technical Information EmulsionSENS Technische Information EmulsionSENS 56925TI1M Rev 00 07 2019...