LB 470RID Level
5 Electric Installation
56925-1BA2 Rev.02, 05/2021
49
4.
Run the cables through the cable glands (Fig. 19, item 6) through the open-
ings of the wall housing and through the counternut cable glands (Fig. 19,
item 7).
5.
Screw the cable glands (Fig. 19, item 6) with the counternut cable glands
(Fig. 19, item 7).
NOTICE
Apply the voltage of the specified and marked (Fig. 19, item 3) range only!
NOTICE
Note the specification relating to Cables, Protective earth, equipotential bond-
ing and EIA-485 (RS-485) in chapter 5.1.
6.
Connect the lines according to assignment (see 2.3 Connection Diagram Ter-
minal Board Master/Master or 2.4 Connection Diagram Terminal Board
Master/Slave in “Technical Information”).
7.
Open the terminal connection (Fig. 19, item 4) with an operating tool (slot-
ted screwdriver) and insert the stripped wire (min. 8 mm) (Fig. 19, item 5).
The terminal connection closes by pulling out the operating tool. The termi-
nal connections are designed for the flexible wires:
- 0.2 mm² … 2.5 mm² or AWG 24 …12 without end sleeve
- 0.25 mm² ... 2.5 mm² with end sleeve without plastic sleeve
- 0.25 mm² ... 1.5 mm² with end sleeve with plastic sleeve.
8.
Plug the network plug into the RJ45 socket (Fig. 19, item 9) (optional).
9.
Tighten all cable glands (Fig. 19, item 6) to ensure optimal sealing and ten-
sion relief until the gasket insert closes between screw down nut and cable.
10.
Check tension relief of all cable glands by pulling the cables smoothly.
The cables must not move. If necessary tighten the cap nuts of the cable
glands.
11.
Slide the subrack into the wall housing and lock it with the square wrench.
NOTICE
The wall enclosure is supplied with blanking elements in all cable glands. It must
be ensured that there are blanking elements in all unused cable glands. Other-
wise, the IP protection is not given.
NOTICE
Only use cable that is suitable for connection to the corresponding terminals may
be used. Detailed specifications can be found in the chapter 5.1.2 Cables and
Lines.
Tip
The connections of the installed devices are already manufactured above the
circuit board in the wall housing. If additional modules (e.g. from other wall
housings) are connected, the terminals appropriate for use on the terminal
board should be used.
The connection was made correctly.
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...