7 Main Menu Device Setup
LB 470RID Level
152
56925-1BA2 Rev.02, 05/2021
Digital Outputs (DO)
Device Setup | Setup | Output | Digital Outputs (DO)
The signals of the digital outputs are switched via potential-free relay contacts. The
contacts are controlled “fail safe”, i.e., in the event of an alarm, the current at the
relay coil drops and the NO contact (normally open) is opened. The wiring diagrams
in the document “Technical Information” show the relay contacts in the de-ener-
gized state.
1
2
3
State DO-1 (ACTIVE / PASSIVE)
State DO-2 (ACTIVE / PASSIVE)
Selection arrow Function DO-2
4
5
6
State DO-3 (ACTIVE / PASSIVE)
Selection arrow Function DO-3
Function selection
Fig. 120 Digital Outputs
The alarm relays 1 and 2 can be assigned to the following functions in the event of
an alarm:
PV Low Alarm
The relay alarms when the value at
Device
Setup | Setup | Alarms | PV alarm settings
is
below the threshold.
PV High Alarm
The relay alarms if the value under
Device
Setup | Setup | Alarms | PV Alarm Settings
is
exceeded.
Detector Temperature Alarm
The relay alarms when values set at
Device
Setup | Setup | Alarms | Det.-Temp. Alarm
function
are exceeded or below the thresh-
old.
XIP Alarm
The relay alarms when detection is acti-
vated at
Device Setup | Setup | Signal
processing | XIP
and interference was de-
tected, so that the measurement was
frozen. See next chapter (Interference Ra-
diation Alarm)
RID Alarm
The relay alarms when detection is acti-
vated at
Device Setup | Setup | Signal
processing | RID
and interference was de-
tected, but the measurement can still
measure in the secondary channel. See next
chapter (Interference Radiation Alarm)
Source Replacement Alarm
The relay alarms when notification at De-
vice Setup | Setup | Signal processing |
Source replacement is activated and inter-
ference is detected.
Measurement Stopped
The relay alarms on during tests or other
states where the measurement is stopped.
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...