7 Main Menu Device Setup
LB 470RID Level
118
56925-1BA2 Rev.02, 05/2021
A new window opens.
5.
Click on the button
<Start>
(Fig. 89, item 5)
.
The measurement starts.
6.
Click
< Ok >
to accept the values.
The level has been adjusted to the process value (see standard display).
IMPORTANT
When executing an Adjust, the measurement parameter set was overwritten. If
the system should be recalibrated, a Recall of the measurement parameter set
to the calibration parameter must be performed. Otherwise the Adjust will be-
come lost.