LB 470RID Level
5 Electric Installation
56925-1BA2 Rev.02, 05/2021
43
5
Electric Installation
5.1
General Instructions
j
DANGER
Danger to life from electric shock!
The installation may only be carried out by a qualified electrician.
Please adhere to the relevant safety regulations.
Open the housing only in a dry environment and for installation, mainte-
nance and servicing.
During installation and servicing on the hardware as well as during wiring
of the detector, the measuring system, connected relay contacts and all in-
puts and outputs must be de-energized.
Connect only devices onto the product that comply with the applicable
safety standards.
In case of an electric shock, carry out first aid measures and immediately call an
emergency service.
NOTICE
Apply the voltage of the specified and marked range only!
The relay of the LB 470RID can only switch low voltages. Please note the speci-
fications in appendix Technical Information.
NOTICE
Installation details and information for intrinsically safe versions of the evalua-
tion unit are listed in a separate safety manual / explosion protection manual
56925BA26.
The power source of 24 V DC version of the product must meet the requirements
of the Low Voltage Directive and be equipped with double or reinforced insula-
tion.
The voltage parameters of all devices connected to the outputs of the product (e.g.
relay circuit, RS-485, current output) must comply with the limit values of the safety
directives for electrical measurement, control, regulation and laboratory devices
(DIN EN 61010-1) and be equipped with double or reinforced insulation. These pro-
tective measures are necessary to avoid the risk of contact with life-threatening
voltages. Changing the installation without precise knowledge of this operating
manual is not permitted.
General important points for installation
Connect the earth conductor.
Ground the housing.
Please observe the information signs on the devices.
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...