3 System Description
LB 470RID Level
30
56925-1BA2 Rev.02, 05/2021
3.5
Measurement Arrangements
The detector and/or the source are rod-shaped for a radiometric level measure-
ment, so as to form a triangular or rectangular useful beam field. The change of
the measurement signal
for different level results from the different sized covering
of the radiation field.
The marking grooves on the detector housing highlight the sensitive area of the
detector. The detector must be mounted on the container so that the desired
measuring range is covered by the sensitive area.
In a cascaded system (with multiple detectors), the lower marking groove of the
1st detector must match the upper of the 2nd detector etc. (Fig. 9).
1
2
3
4
5
6
7
Point source / shield
Mounting base with source
Level
Radiation field
Upper marking groove
Rod detector
Lower marking groove
Fig. 8
Schema point source - rod detector
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...