7 Main Menu Device Setup
LB 470RID Level
102
56925-1BA2 Rev.02, 05/2021
Detector settings: Detector Service | Reset Detector
Device Setup | Setup | Sensors | [NAME DETECTOR] | Detector Service | Reset Detec-
tor
In the window “Reset Detector”, the detector can be restarted and be reset to the
factory settings.
1
2
3
4
Button Restart
Button Factory settings
Window "Warning: Reboot"
Window "Warning: Factory Reset"
Fig. 72
Reset Detector
IMPORTANT
The measurement is interrupted during a restart!
1.
To restart the Detector, click the button
<Reboot>
(Fig. 72, item 1).
A window with a warning “Reboot” (Fig. 72, item 3) opens.
2.
Click on
<Yes>
to confirm.
The device is restarted.
IMPORTANT
All custom configuration settings will be lost with a reset to factory settings!
1.
To reset the detector to the factory settings, click the button
<Factory reset>
(Fig. 72, item 2).
A window with the warning Factory settings (Fig. 72, item 4) opens.
2.
Click on
<Yes>
to confirm.
The device is reset to factory settings and restarts.
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...