LB 470RID Level
7 Main Menu Device Setup
56925-1BA2 Rev.02, 05/2021
103
7.3.3
Calibration
Device Setup | Setup | Calibration
The Calibration menu is used for the selection of the calibration and calculation
method and for adaption of the measurement system to the respective environ-
mental conditions, the actual radiation activity and the adaptation of the
background radiation (background level).
NOTICE
Material damage to the device or the system!
Errors in calibration or in the parameter setting can lead to incorrect measure-
ment results. This may possibly lead to loss of production or to damage in the
system.
We encourage you to have the calibration and commissioning performed by
Berthold service.
IMPORTANT
The RID function must be activated in the menu
Device Setup | Setup | Signal
processing | RID
.
Interference Radiation Suppression using the LB 470RID operates only with
detectors having plastic scintillators; these are especially susceptible to inter-
ference radiation on account of their large volumes.
Fig. 73
Menu “Calibration”
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...