7 Main Menu Device Setup
LB 470RID Level
138
56925-1BA2 Rev.02, 05/2021
Signal processing: RID
Device Setup | Setup | Signal Condition | RID
1
2
3
4
5
Checkbox “Enabled”
Input field “PV Threshold”
Button
<Expert Settings>
Input field “RID Time Constant”
Input field “RID Limit”
6
7
8
9
10
Input field “Min. Weight”
Input field ”Max. Weight”
Input field ”Daily Weight Reduction”
Button
<Expert-Settings>
Button
<Reset RID Table>
Fig. 109 Signal Condition (RID)
NOTICE
The following parameters should be changed only by Berthold employees, or in
consultation with Berthold.
Explanation "Weight"
The RID table consists of 100 points. Through these points, a characteristic curve is
formed from the function between the secondary channel and the measurement
channel. The weight describes how strongly the respective point affects the char-
acteristic curve. The higher the weight the stronger the characteristic is forced by
the respective point.
5
Radiation Interference Discrimination
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...