LB 470RID Level
7 Main Menu Device Setup
56925-1BA2 Rev.02, 05/2021
109
Calibration Settings: Table (linear calibration type)
Device Setup | Setup | Calibration | Calibration Settings
1
2
3
4
5
6
7
Column count rate [%]
Column count rate [cps]
Column count rate [rid]
Button
<Edit>
Button Add measuring point
< + >
Button Delete point
< - >
Date of the last editing
8
9
10
11
12
13
Count of the calibration points
Read-In-Time [s]
Count rate RID [cps]
Count rate [cps]
Process value [%]
Measuring process (determine new
calibration point)
Fig. 80
Calibration parameters (Table linear calibration type)
NOTICE
To avoid calibration errors, it must be ensured that during calibration and com-
missioning of the measuring equipment no welding tests are being done on the
site, even at long distance.
Add new calibration point
1.
Click on the button
< + >
(Fig. 80, item 5).
The window “New calibration point” (Fig. 80, item 9-12) opens.
2.
Click on the input field “Process value” (Fig. 80, item 12) to open the input
field.
3.
Specify the current, actual level in percent and confirm with the Enter key.
4.
Click on the input field “Read-In-Time” (Fig. 80, item 9) to open the input
field.
5.
Specify a read-in time and confirm with the Enter key.
6.
Click on the button
<Start>.
The Window “New calibration Point” (Fig. 80, item 13) opens and the count
rate is determined.
7.
Click on
<Ok>
to confirm the new calibration point.
The new calibration point is recorded in the table.
Содержание DUO XPERT LB 470RID
Страница 2: ......
Страница 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Страница 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Страница 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Страница 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Страница 199: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 04 09 2019...
Страница 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...
Страница 235: ...LB 470RID Level LB 470RID F llstand Technical Information Technische Information 56925TI1L Rev 04 05 2021 28...
Страница 236: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 29 56925TI1L Rev 04 05 2021...
Страница 238: ...Technical Information Technische Information LB 470RID Level LB 470RID F llstand 31 56925TI1L Rev 04 05 2021...