6. Validation using 3H- und 14C-Test sources
FlowStar2 LB 514
52
History
Pressing
<History>
from the validation menu will lead to the test cell se-
lection.
The desired cell can be selected to display their results. After selecting the
cell the cell parameters as well as the previously generated results are
displayed:
The date of the validation, the measurement value, the nominal value at
this date (half live corrected) and the deviation from the nominal value is
shown in the table. The entry above the line show the values from the initial
validation.
Parameters
Pressing
<Parameters>
from the validation menu will lead to the parame-
ters page for the validation.
Prerun Time
Defines the waiting time before starting the validation measure-
ment. This prerun time is used to calm down the test cell which
VALIDATION HISTORY
1 / 1
Test H-3...
Test C-14...
Test BG...
Test BGO...
===========Last / next Validation========================
Last Validation on:
05. Jan. 2010
Next Validation on:
05. Apr. 2010
Esc
VALIDATION HISTORY
1 / 1
Ident Number:
49926
Serial Number:
52003
Cell Type:
C-14
Scintillator Type:
Test C-14
Date/Time
Value
Nominal Dev.
[cpm]
[cpm]
09.04.09 15:06
69220
68959 0.6
07.01.10 17:32
69012
68959 0.5
09.12.09 14:48
69180
68959 0.5
08.09.09 12:24
69235
68959 0.6
12.07.09 12:36
69197
68959 0.6
Esc
VALIDATION PARAMETERS
1 / 1
Prerun Time:
00:00:00
Measurement Time:
00:05:00
Tolerance Threshold [%]:
3.000
Rand. Error f. Background[%]
1.000
Validation Cycle [d]:
90
Validation Mode strict:
x
Last Validation on:
05. Jan. 2010
Next Valation on:
05. Apr. 2010