6. Validation using 3H- und 14C-Test sources
FlowStar2 LB 514
50
6.1 Checking the FlowStar² with test sources
Remove cell.
Install test source (
3
H or
14
C) in place of the measuring cell. Please
make sure that the cell and the test source will not see any light! This
will cut down the decay period.
Start the validation measurement (Validation | Measurement | <Start>)
and wait until the measurement is finished.In case of a passed valida-
tion the measured values can be saved in the history. If the validation
failed check the instrument and repeat the measurement.
Procedure and parameter entry
Select
Validation
from the main menu. This entry is only visible if you are
logged in as supervisor. The following menu appears:
Measurement
Validation measurements are handled through this menu.
History
Earlier, successful validation measurements can be revied in the
Validation history.
Parameters
Validation parameters such as tolerance levels and measuring
time are defined in this menu.
Measurement
When selecting
Measurement
from the validation menu the window to per-
form validation measurements pops up.
MEASUREMENT
Parameters
HISTORY