FlowStar2 LB 514
7. Startup of the HPLC System
65
In these dialog box you can select what kind of extensions will be
installed.
The USB device driver is essential to communicate to the instru-
ment and should not be disabled.
Select the Chromeleon™ fea-
ture only, when you intend to control the FlowStar² using Chro-
meleon® software. Select the correct version of your Chro-
meleon® installation in order to install the drivers properly. Instru-
ment templates are only available for 6.x versions.
For general purpose control or RadioStar just the FlowStar²
driver is required.
The installation starts after clicking on <next>.
After the installation is finished a summary page will appear.
Since the controller uses the TCP/IP protocol for communication
a firewall warning message will show up when the controller
starts up the first time. To guarantee a proper function of the con-
troller communication must be allowed. If there will be no permis-
sion the controller cannot work properly. In this case the windows
firewall must be set to allow “flwCtrl.exe” TCP/IP access.
When connecting the FlowStar² to the PC the first time a mes-
sage will show up that a new device was found. In the upcoming
dialog box select “automatic´installation” to install the drivers who
have been copied by the FlowStar² installation software before.
Please make sure that the driver was installed before con-
necting the instrument to the PC to avoid problems with the
driver installation.