9. Parameter Entries
FlowStar2 LB 514
88
Time window
Active from ... to ...
Enter the time for enabling the fraction collector.
Threshold value
Only if you have selected
Threshold
in the
controlled by
field. Enter
the low and high threshold value.
Interval
Only if you have selected
Threshold
in the
controlled by
field. Enter the
interval time in seconds.
Waste valve parameters
WASTE VALVE SETTINGS
1 / 1
Delay Time [s]:
1.0
On-Threshold:
100.0
Off-Threshold:
0.0
Level Threshold
Figure 32: Waste valve parameter
Delay time [s]
Time delay from the cell output to the waste valve. This time delay de-
pents on the specific flow rate and the length of the tubing used and must
be tested individually.
Threshold value
Enter the low and high threshold value for the waste valve control. Please
keep in mind that the switch-off threshold must be smaller than the
switch-on threshold.