9. Parameter Entries
FlowStar2 LB 514
80
9.2 Nuclide Parameters
The
FlowStar²
software includes a nuclide table on the
Nuclide
menu
which contains the major nuclides for the HPLC methods including half-life
period, energy thresholds and the measuring cells that can be used. The
parameters of these nuclides cannot be changed by the user.
You can also define your own nuclides or nuclide mixtures with user-spe-
cific parameters in the nuclide table.
Select
Menu | Nuclide
on the measurement menu.
NUCLIDES
1 / 3
H-3...
C-14
P-3
...
2...
S-35...
P-33...
Cr-51...
I-125...
Tc-99m...
I-129...
PET...
In-111...
Def 11...
NUCLIDE PARAMETERS
1 / 1
Name:
H-3
Activeated:
x
Half Life Time [h]:
107748
Allowed Cells...
Lower Channel:
20
Upper Channel:
1023
Energy Window
Figure 25: Nuclide table (left); parameters of a selected nuclide (right)
The 1
st
page of the
nuclide table
lists the supplied nuclides; on the follow-
ing pages you can define your own nuclides. These have the default name
Def 11
,
Def 12
etc. or
USR-0
,
USR-1
etc.
Editing a user-defined nuclide
Highlight the desired nuclide (Defxx or USR-x), and push the
button.
The page with the respective nuclide parameters is displayed.
Name
Enter a name for the nuclide (max. 6 characters).
Enabled
Only if the checkbox of a nuclide is ticked (x), it can be selected for a
method.
Half-life [h]
Information on the half-life period in hours.
Permitted cells...
If you select this item, a list of available measuring cells is displayed. Tick
the measuring cells (x) that are allowed for the respective nuclide. Thus,
the nuclides allowed for a method are already selected when inserting a
measuring cell.
Energy window
Gain high/ low
The gain setting for the amplifier must be set here. Depending on the
expected signal the gain must be either low (x1) or high (x10).
Low / High chan-
nel
threshold
Define the energy window for the respective nuclide by entering a high
and low threshold. If 2 windows are required for a nuclide, define new
nuclides accordingly (see below).