NOTICE
The foot pump pedal is located on the left side of the patient (with
table in normal orientation) near the head end of the base. Using
the foot pump pedal deactivates the hand pendant and activates
the auxiliary hand pendant.
1. Keeping it straight, pull out the foot pump pedal to the first end stop.
2. Turn pedal counterclockwise.
3. Pull out the foot pump pedal to the second end stop.
¡
Foot pump is now activated and the main hand pendant id de-acti-
vated.
1. Push the pedal back into the table until the end reaches the base
cover.
2. Push the pedal further into the table until it snaps into position.
¡
The foot pump is deactivated and the hand pendant is activated.
Activating the foot pump
Deactivating the foot pump
Emergency release
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Malfunctions
98