9.4.1 Auxiliary hand pendant D 850
The following prerequisites must be met:
¡
The foot pump is activated (see section 9.3).
¡
The Auxiliary hand pendant is activated, the AUX PENDANT indicator
light on the hand pendant is yellow.
1. Take the Auxiliary hand pendant from its storage box (head end) (2).
2. To tilt the table, press and hold the desired function key on the emer-
gency hand pendant (1).
3. Simultaneously operate the foot pump pedal (2) with your foot.
ð
The OR table performs the desired function.
4. Switch off the Auxiliary hand pendant.
ð
The AUX PENDANT light on the Auxiliary hand pendant goes out.
5. Return the Auxiliary hand pendant to its storage box (2) after use.
NOTICE
The following functions will not work with the Auxiliary Hand
Pendent: Slide, Level, Chair and Flex.
9.4.2 Auxiliary hand pendant D 820, D 760
The following prerequisites must be met:
¡
The foot pump is activated
¡
The Auxiliary hand pendant is activated.
1. To tilt the table, press the desired function key on the Auxiliary hand
pendant (1).
2. Operate the foot pump pedal with your foot.
– The OR table performs the desired function.
3. Switch off the Auxiliary hand pendant / return foot pump to its starting
position.