A maximum relative humidity of 95% is ensured.
¡
Humidity is less than or equal to 95%
¡
The OR table is not exposed to an explosive atmosphere.
¡
The OR table is not exposed to a corrosive atmosphere.
¡
The OR table is not exposed to direct sunlight.
¡
Ambient heat sources cannot add external heat.
¡
There are no sources of dust.
¡
Fire prevention measures have been taken.
¡
The OR table is not exposed to vibration/oscillation.
¡
The flooring is resistant to solvents, impervious to liquids, antistatic and
easy to clean.
¡
There are no machines nearby that would cause electrical or electro-
magnetic interference. See section 11.3.
3.2 Unpacking and installing
Protective equipment:
¡
Protective clothing
¡
Protective gloves
¡
Safety boots
Special tools:
¡
Spirit level
¡
Protractor
¡
Measuring tape
¡
Wrench 11 mm
¡
Wrench 14 mm
¡
Wire cutters
¡
Flat screwdriver
¡
Pliers