Materials:
¡
Quaternary ammonium compounds or phenol-containing disinfec-
tants
CAUTION
Risk of operator injury from compromised conductivity of patient
pads due to damage or soiling!
Conductivity may be compromised if pads are not properly
cleaned and maintained.
1. Replace damaged pads immediately.
NOTICE
Material damage from disinfectants such as lodophor!
Using disinfectants such as lodophor causes spotting on pads
and reduces their service life.
1. Use only quaternary ammonium compounds or phenol-
containing disinfectants for cleaning.
NOTICE
Material damage from incorrect cleaning agents!
Using the wrong cleaning agents can damage the pads.
1. Do not use any solvents, such as thinners or acetone, to
clean pads.
2. Do not treat pads in sterilizers.
3. Do not use any sharp, pointed or abrasive objects for
cleaning.
Cleaning conductive
objects like pads and
straps
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Operation