NOTICE
Maintenance tasks are explained in the separate service manual.
On request, BERCHTOLD provides circuit diagrams, parts lists,
descriptions, and other information needed for repairing the OR
table.
Interval
Maintenance task
Every 6 months
Check operation of hydraulic sys-
tem
Annually
General maintenance
After each use
Clean the OR table (see “Cleaning
the OR table”)
Charge batteries (see “Charging
the batteries”)
Maintenance item
Maintenance interval
Check:
Electrical safety (visual)
Monthly
Operation of hydraulic system
Every 6 months
Ground continuity/leakage current Every 6 months
Flexible hydraulic lines
Annually
Condition of hydraulic fluid
Annually or if a leak is suspected or
if hydraulic components were re-
placed.
Fasteners and restraining clamps
Annually
Clamps and accessories
Monthly (accessories can be
changed often)
Screws on the longitudinal slide
drive belt
Annually
Structural integrity of the table
Annually
Spare part:
Batteries
Every 3 years or if the charge/
discharge cycle
drops below 5 – 8 hours.
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Maintenance
94