See Notice on page 6 for single press information
– The indicator light flashes yellow after the first press and the un-
locking process is confirmed within 3 seconds after the second
press.
– The base locking cylinders move up and unlock the OR table. The
UNLOCK indicator light turns yellow when the process is finished.
– The OR table can now be moved.
1. Lock activate light
2. Lock/unlock button
3. Unlock activate light
4.7 Setting reverse orientation
Personnel:
WARNING
Risk of injury due to incorrect patient positioning and unintended
movement of the OR table!
Reverse orientation reverses some control functions. The opera‐
tor must be aware of these functions and the ori-entation of the
OR table. Failure to do this can result in movement of the wrong
section and serious injury. See NOTICE on page 33
1. Do not lean on the different sections or use them for sup-
port.
2. Comply with specifications provided in the technical da-
ta on patient weight, size, and positioning. (“Patient po-
sitioning” section)
3. The patient's pelvis must not extend over the end of the
seat section.
4. In reverse orientation, the maximum patient weight must
not exceed 500 lb. (227 kg).
5. Carefully monitor all OPERON® table movements and
stop immediately when in doubt.